![Rohde & Schwarz FSW Series User Manual Download Page 106](http://html.mh-extra.com/html/rohde-and-schwarz/fsw-series/fsw-series_user-manual_1477844106.webp)
Configuration
R&S
®
FSW-K7
106
User Manual 1173.9240.02 ─ 17
Sweep Points
Defines the number of measured values to be collected during one sweep.
All values from 101 to 100001 can be set. The default value is 1001 sweep points.
Remote command:
Sweep/Average Count
Defines the number of sweeps to be performed in the single sweep mode. Values from
0 to 200000 are allowed. If the values 0 or 1 are set, one sweep is performed.
The sweep count is applied to all the traces in all diagrams.
If the trace modes "Average", "Max Hold" or "Min Hold" are set, this value also deter-
mines the number of averaging or maximum search procedures.
In continuous sweep mode, if sweep count = 0 (default), averaging is performed over
10 sweeps. For sweep count =1, no averaging, maxhold or minhold operations are per-
formed.
Remote command:
5.7 Demodulation Display
The demodulated signal can be displayed using various evaluation methods. All evalu-
ation methods available for the Analog Demodulation application are displayed in the
evaluation bar in SmartGrid mode when you do one of the following:
●
Select the
"SmartGrid" icon from the toolbar.
●
Select the "Demod/Display" button in the "Overview".
●
Press the MEAS key.
●
Select the "Display Config" softkey in the main "Analog Demod" menu.
Up to six evaluation methods can be displayed simultaneously in separate windows.
The Analog Demodulation evaluation methods are described in
For details on working with the SmartGrid see the R&S
FSW Getting Started manual.
5.8 Demodulation
Demodulation parameters can be configured in the "Demodulation Settings" dialog box
which is displayed when you select the "Demod Settings" button in the "Overview".
Demodulation Display