Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
295
User Manual 1173.9240.02 ─ 17
Example:
CALC:FEED 'XTIM:PM:TDOM'
Switches on the PM time domain result display.
DISP:TRAC ON
Switches on the trace.
CALC:MARK:FUNC:ADEM:PM? PPE
Queries the peak value of the demodulated PM trace.
Usage:
Query only
Manual operation:
See
CALCulate:MARKer:FUNCtion:ADEMod:CARRier[:RESult]?
This command queries the carrier power, which is determined from the Clr/Write data.
Return values:
<CPower>
Power of the carrier without modulation in dBm.
Usage:
Query only
CALCulate:MARKer:FUNCtion:ADEMod:FERRor[:RESult<t>]?
This command queries the carrier offset (= frequency error) for FM and PM demodula-
tion. The carrier offset is determined from the current measurement data (CLR/
WRITE). The modulation is removed using low pass filtering.
The offset thus determined differs from that calculated in the
command which uses averaging to determine the frequency deviation.
Return values:
<CarrOffset>
The deviation of the calculated carrier frequency to the ideal car-
rier frequency in Hz.
Usage:
Query only
CALCulate<n>:MARKer:FUNCtion:ADEMod:SINad:RESult<t>?
This command queries the result of the signal-to-noise-and-distortion (SINAD) mea-
surement in the specified window for the specified trace.
Note that this value is only calculated if an AF Spectrum window is displayed.
Parameters:
<SINAD>
The signal-to-noise-and-distortion ratio in dB.
Usage:
Query only
CALCulate<n>:MARKer:FUNCtion:ADEMod:THD:RESult<t>?
This command queries the result of the total harmonic distortion (THD) measurement
in the specified window.
Note that this value is only calculated if an AF Spectrum window is displayed.
Retrieving Results