Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
302
User Manual 1173.9240.02 ─ 17
If the marker has been used as a delta marker, the command turns it into a normal
marker.
Parameters:
<Position>
Numeric value that defines the marker position on the x-axis.
The unit is either Hz (frequency domain) or s (time domain) or
dB (statistics).
Range:
The range depends on the current x-axis range.
Example:
CALC:MARK2:X 1.7MHz
Positions marker 2 to frequency 1.7 MHz.
Manual operation:
See
See
See
CALCulate<n>:MARKer<m>:Y?
This command queries the position of a marker on the y-axis.
If necessary, the command activates the marker first.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps. See also
If the analog demodulator (option Analog Demodulation, R&S
FSW–K7) is activated,
the query result is output in the following units in the specified window:
Result display
Output unit
AM
%
FM
Hz
PM
RF
dB (Range Log or Range Linear %)
% (Range Linear dB)
Return values:
<Result>
Result at the marker position.
Example:
INIT:CONT OFF
Switches to single measurement mode.
CALC:MARK2 ON
Switches marker 2.
INIT;*WAI
Starts a measurement and waits for the end.
CALC:MARK2:Y?
Outputs the measured value of marker 2.
Usage:
Query only
Analyzing Results