Configuration
R&S
®
FSW-K7
97
User Manual 1173.9240.02 ─ 17
For more information on the the Digital Baseband Interface or the Analog Baseband
Interface see the R&S
FSW I/Q Analyzer and I/Q Input User Manual.
Remote command:
TRIG:SOUR BBP
, see
Digital I/Q ← Trigger Source
For applications that process I/Q data, such as the I/Q Analyzer or optional applica-
tions, and only if the Digital Baseband Interface (R&S
FSW-B17) is available:
Defines triggering of the measurement directly via the LVDS connector. In the selection
list you must specify which general purpose bit (GP0 to GP5) will provide the trigger
data.
Note:
If the Digital I/Q enhanced mode is used, i.e. the connected device supports transfer
rates up to 200 Msps, only the general purpose bits GP0 and GP1 are available as a
Digital I/Q trigger source.
The following table describes the assignment of the general purpose bits to the LVDS
connector pins.
(For details on the LVDS connector see the R&S
FSW I/Q Analyzer User Manual.)
Table 5-2: Assignment of general purpose bits to LVDS connector pins
Bit
LVDS pin
GP0
SDATA4_P - Trigger1
GP1
SDATA4_P - Trigger2
GP2
*)
SDATA0_P - Reserve1
GP3
*)
SDATA4_P - Reserve2
GP4
*)
SDATA0_P - Marker1
GP5
*)
SDATA4_P - Marker2
*)
: not available for Digital I/Q enhanced mode
Remote command:
TRIG:SOUR GP0
, see
FM / AM / PM / RF (Offline) ← Trigger Source
Triggers when the demodulated input signal exceeds the trigger level.
Remote command:
Time ← Trigger Source
Triggers in a specified repetition interval.
Remote command:
TRIG:SOUR TIME
, see
Trigger Configuration