Continuous Sweep/RUN CONT
After triggering, starts the sweep and repeats it continuously until stopped. This is the
default setting.
While the measurement is running, the "Continuous Sweep" softkey and the RUN
CONT key are highlighted. The running measurement can be aborted by selecting the
highlighted softkey or key again. The results are not deleted until a new measurement
is started.
Note:
Sequencer. If the Sequencer is active, the "Continuous Sweep" softkey only
controls the sweep mode for the currently selected channel; however, the sweep mode
only has an effect the next time the Sequencer activates that channel, and only for a
channel-defined sequence. In this case, a channel in continuous sweep mode is swept
repeatedly.
If the Sequencer is active in MSRT mode, the "Continuous Sweep" function does not
start data capturing; it merely has an effect on trace averaging over multiple sequen-
ces. In this case, trace averaging is performed.
Furthermore, the RUN CONT key controls the Sequencer, not individual sweeps. RUN
CONT starts the Sequencer in continuous mode.
For details on the Sequencer, see the R&S
FSW User Manual.
Remote command:
Single Sweep/ RUN SINGLE
After triggering, starts the number of sweeps set in "Sweep Count". The measurement
stops after the defined number of sweeps has been performed.
While the measurement is running, the "Single Sweep" softkey and the RUN SINGLE
key are highlighted. The running measurement can be aborted by selecting the high-
lighted softkey or key again.
Data Acquisition