Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
234
User Manual 1173.9240.02 ─ 17
Remote commands exclusive to amplitude configuration:
CALCulate<n>:MARKer<m>:FUNCtion:REFerence
........................................................... 234
........................................................................................... 234
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:RLEVel
............................................................ 234
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:RLEVel:OFFSet
................................................235
CALCulate<n>:MARKer<m>:FUNCtion:REFerence
This command matches the reference level to the power level of a marker.
If you use the command in combination with a delta marker, that delta marker is turned
into a normal marker.
Example:
CALC:MARK2:FUNC:REF
Sets the reference level to the level of marker 2.
Usage:
Event
CALCulate<n>:UNIT:POWer
<Unit>
This command selects the unit of the y-axis.
The unit applies to all measurement windows.
Parameters:
<Unit>
DBM | V | A | W | DBPW | WATT | DBUV | DBMV | VOLT |
DBUA | AMPere
*RST:
dBm
Example:
CALC:UNIT:POW DBM
Sets the power unit to dBm.
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:RLEVel
<ReferenceLevel>
This command defines the reference level.
With a reference level offset
≠
0, the value range of the reference level is modified by
the offset.
Parameters:
<ReferenceLevel>
The unit is variable.
Range:
see datasheet
*RST:
0 dBm
Example:
DISP:TRAC:Y:RLEV -60dBm
Usage:
SCPI confirmed
Manual operation:
See
Configuring the Measurement