Configuration
R&S
®
FSW-K7
72
User Manual 1173.9240.02 ─ 17
"I + jQ"
The input signal is filtered and resampled to the sample rate of the
application.
Two inputs are required for a complex signal, one for the in-phase
component, and one for the quadrature component.
"I Only / Low IF I"
The input signal at the BASEBAND INPUT I connector is filtered and
resampled to the sample rate of the application.
If the center frequency is set to 0
Hz, the real baseband signal is dis-
played without down-conversion (
Real Baseband I
).
If a center frequency greater than 0
Hz is set, the input signal is
down-converted with the center frequency (
Low IF I
).
"Q Only / Low IF Q"
The input signal at the BASEBAND INPUT Q connector is filtered and
resampled to the sample rate of the application.
If the center frequency is set to 0
Hz, the real baseband signal is dis-
played without down-conversion (
Real Baseband Q
).
If a center frequency greater than 0
Hz is set, the input signal is
down-converted with the center frequency (
Low IF Q
).
Remote command:
Input configuration
Defines whether the input is provided as a differential signal via all 4 Analog Baseband
connectors or as a plain I/Q signal via 2 simple-ended lines.
Note:
Both single-ended and differential probes are supported as input; however, since
only one connector is occupied by a probe, the "Single-ended" setting must be used for
all probes.
"Differential"
I, Q and inverse I,Q data
"Single Ended" I, Q data only
Remote command:
Center Frequency
Defines the center frequency for analog baseband input.
For real-type baseband input (I or Q only), the center frequency is always 0
Hz.
Note:
If the analysis bandwidth to either side of the defined center frequency exceeds
the minimum frequency (0
Hz) or the maximum frequency (40
MHz/80
MHz), an error
is displayed. In this case, adjust the center frequency or the analysis bandwidth.
Remote command:
5.4.1.5
Probe Settings
Probes are configured in a separate tab on the "Input" dialog box which is displayed
when you select the INPUT/OUTPUT key and then "Input Source Config".
Input and Frontend Settings