Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
268
User Manual 1173.9240.02 ─ 17
Example:
ADJ:CONF:DUR:MODE MAN
Selects manual definition of the measurement length.
ADJ:CONF:LEV:DUR 5ms
Length of the measurement is 5 ms.
Manual operation:
See
"Changing the Automatic Measurement Time (Meastime
[SENSe:]ADJust:CONFigure:DURation:MODE
<Mode>
In order to determine the ideal reference level, the R&S
FSW performs a measurement
on the current input data. This command selects the way the R&S
FSW determines the
length of the measurement .
Parameters:
<Mode>
AUTO
The R&S
FSW determines the measurement length automati-
cally according to the current input data.
MANual
The R&S
FSW uses the measurement length defined by
*RST:
AUTO
Manual operation:
See
"Resetting the Automatic Measurement Time (Meastime
See
"Changing the Automatic Measurement Time (Meastime
[SENSe:]ADJust:FREQuency
This command sets the center frequency to the highest signal level in the current fre-
quency range.
Example:
ADJ:FREQ
Usage:
Event
Manual operation:
See
"Adjusting the Center Frequency Automatically (Auto Freq)"
[SENSe:]ADJust:CONFigure:HYSTeresis:LOWer
<Threshold>
When the reference level is adjusted automatically using the
on page 269 command, the internal attenuators and the preamplifier are also
adjusted. In order to avoid frequent adaptation due to small changes in the input signal,
you can define a hysteresis. This setting defines a lower threshold the signal must fall
below (compared to the last measurement) before the reference level is adapted auto-
matically.
Configuring the Measurement