Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
267
User Manual 1173.9240.02 ─ 17
11.4.9 Adjusting Settings Automatically
The following remote commands are required to adjust settings automatically in a
remote environment. The tasks for manual operation are described in
MSRA/MSRT operating mode
In MSRA/MSRT operating mode, settings related to data acquisition cannot be adjus-
ted for Analog Demodulation applications.
[SENSe:]ADJust:CONFigure:DURation
............................................................................ 267
[SENSe:]ADJust:CONFigure:DURation:MODE
..................................................................268
..........................................................................................268
[SENSe:]ADJust:CONFigure:HYSTeresis:LOWer
..............................................................268
[SENSe:]ADJust:CONFigure:HYSTeresis:UPPer
...............................................................269
[SENSe:]ADJust:SCALe:Y:AUTO[:CONTinuous]
............................................................... 269
[SENSe:]ADJust:ALL
This command initiates a measurement to determine and set the ideal settings for the
current task automatically (only once for the current measurement).
This includes:
●
Center frequency
●
Reference level
●
Scaling
Example:
ADJ:ALL
Usage:
Event
Manual operation:
See
"Adjusting all Determinable Settings Automatically (Auto
[SENSe:]ADJust:CONFigure:DURation
<Duration>
In order to determine the ideal reference level, the R&S
FSW performs a measurement
on the current input data. This command defines the length of the measurement if
is set to
MANual
.
Parameters:
<Duration>
Numeric value in seconds
Range:
0.001 to 16000.0
*RST:
0.001
Default unit: s