Configuration
R&S
®
FSW-K7
122
User Manual 1173.9240.02 ─ 17
How to provide trigger signals as output is described in detail in the R&S
FSW User
Manual.
Digital I/Q output is not available for Analog Demodulation measurements.
Output settings can be configured via the INPUT/OUTPUT key or in the "Outputs" dia-
log box.
.......................................................................................123
....................................................................................... 123
Noise Source
Switches the supply voltage for an external noise source on or off.
External noise sources are useful when you are measuring power levels that fall below
the noise floor of the R&S
FSW itself, for example when measuring the noise level of a
DUT.
Remote command:
Trigger 2/3
Defines the usage of the variable TRIGGER INPUT/OUTPUT connectors, where:
"Trigger 2": TRIGGER INPUT/OUTPUT connector on the front panel
"Trigger 3": TRIGGER 3 INPUT/ OUTPUT connector on the rear panel
(Trigger 1 is INPUT only.)
Note:
Providing trigger signals as output is described in detail in the R&S
FSW User
Manual.
Output Settings