Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
280
User Manual 1173.9240.02 ─ 17
INITiate:SEQuencer:ABORt
This command stops the currently active sequence of measurements. The Sequencer
itself is not deactivated, so you can start a new sequence immediately using
To deactivate the Sequencer use
Usage:
Event
INITiate:SEQuencer:IMMediate
This command starts a new sequence of measurements by the Sequencer. Its effect is
similar to the
command used for a single measurement.
Before this command can be executed, the Sequencer must be activated (see
Example:
SYST:SEQ ON
Activates the Sequencer.
INIT:SEQ:MODE SING
Sets single sequence mode so each active measurement will be
performed once.
INIT:SEQ:IMM
Starts the sequential measurements.
Usage:
Event
INITiate:SEQuencer:MODE
<Mode>
This command selects the way the R&S
FSW application performs measurements
sequentially.
Before this command can be executed, the Sequencer must be activated (see
A detailed programming example is provided in the "Operating Modes" chapter in the
R&S
FSW User Manual.
Note:
In order to synchronize to the end of a sequential measurement using *OPC,
*OPC? or *WAI you must use
SINGle
Sequence mode.
For details on synchronization see the "Remote Basics" chapter in the R&S
FSW User
Manual.