![Rohde & Schwarz FSW Series User Manual Download Page 254](http://html.mh-extra.com/html/rohde-and-schwarz/fsw-series/fsw-series_user-manual_1477844254.webp)
Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
254
User Manual 1173.9240.02 ─ 17
OUTPut:TRIGger<port>:OTYPe
<OutputType>
This command selects the type of signal generated at the trigger output.
Note
: For offline AF or RF triggers, no output signal is provided.
Suffix:
<port>
.
Selects the trigger port to which the output is sent.
2 = trigger port 2 (front)
3 = trigger port 3 (rear)
Parameters:
<OutputType>
DEVice
Sends a trigger signal when the R&S
FSW has triggered inter-
nally.
TARMed
Sends a trigger signal when the trigger is armed and ready for
an external trigger event.
UDEFined
Sends a user defined trigger signal. For more information see
*RST:
DEVice
Manual operation:
See
OUTPut:TRIGger<port>:PULSe:IMMediate
This command generates a pulse at the trigger output.
Suffix:
<port>
.
Selects the trigger port to which the output is sent.
2 = trigger port 2 (front)
3 = trigger port 3 (rear)
Usage:
Event
Manual operation:
See
OUTPut:TRIGger<port>:PULSe:LENGth
<Length>
This command defines the length of the pulse generated at the trigger output.
Suffix:
<port>
.
Selects the trigger port to which the output is sent.
2 = trigger port 2 (front)
3 = trigger port 3 (rear)
Parameters:
<Length>
Pulse length in seconds.
Manual operation:
See
Configuring the Measurement