Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
326
User Manual 1173.9240.02 ─ 17
Manual operation:
See
CALCulate:LIMit<k>:LOWer:SPACing
<InterpolType>
This command selects linear or logarithmic interpolation for the calculation of a lower
limit line from one horizontal point to the next.
Parameters:
<InterpolType>
LINear | LOGarithmic
*RST:
LIN
Manual operation:
See
See
CALCulate:LIMit<k>:LOWer:STATe
<State>
This command turns a lower limit line on and off.
Before you can use the command, you have to select a limit line with
Parameters:
<State>
ON | OFF
*RST:
OFF
Usage:
SCPI confirmed
Manual operation:
See
CALCulate:LIMit<k>:LOWer:THReshold
<Threshold>
This command defines a threshold for relative limit lines.
The R&S
FSW uses the threshold for the limit check, if the limit line violates the thresh-
old.
Parameters:
<Threshold>
Numeric value.
The unit depends on
on page 327.
*RST:
-200 dBm
Manual operation:
See
CALCulate:LIMit<k>:NAME
<Name>
This command selects a limit line that already exists or defines a name for a new limit
line.
Parameters:
<Name>
String containing the limit line name.
*RST:
REM1 to REM8 for lines 1 to 8
Analyzing Results