Configuration
R&S
®
FSW-K7
75
User Manual 1173.9240.02 ─ 17
TTL Handshake
If available for the specified generator type, this option activates TTL synchronization
via handshake for GPIB connections.
Using the TTL interface allows for considerably higher measurement rates than pure
GPIB control, because the frequency stepping of the R&S
FSW is directly coupled with
the frequency stepping of the generator.
For more information on TTL synchronization see
For an overview of which generators support TTL synchronization see
"Overview of Generators Supported by the R&S
Remote command:
GPIB Address / TCP/IP Address
For LAN connections: TCP/IP address of the signal generator
For GPIB connections: GPIB address of the signal generator.
Remote command:
Reference
Selects the internal R&S
FSW or an external frequency reference to synchronize the
R&S
FSW with the generator (default: internal).
Remote command:
Edit Generator Setup File
Displays the setup file for the currently selected
an editor.
Although the existing setup files are displayed in read-only mode in the editor, they can
be saved under a different name (using "File > SaveAs").
Be careful, however, to adhere to the required syntax and commands. Errors will only
be detected and displayed when you try to use the new generator (see also
ter 4.7.4.8, "Displayed Information and Errors"
For details see
chapter 4.7.4.3, "Generator Setup Files"
Frequency Min. / Frequency Max.
For reference only: Lower and upper frequency limit for the generator.
Level Min. / Level Max.
For reference only: Lower and upper power limit for the generator.
Measurement Settings
The measurement settings for external generator control are configured in the "Mea-
surement Configuration" subtab of the "External Generator" tab.
Input and Frontend Settings