Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
176
User Manual 1173.9240.02 ─ 17
Suffix
Value range
Description
<n>
1..6
Window or Evaluation
<t>
1..6
Trace
11.3 Activating Analog Demodulation Measurements
Analog demodulation measurements require a special application on the R&S
FSW.
The measurement is started immediately with the default settings.
...................................................................................... 176
............................................................................................ 176
........................................................................................ 177
SYSTem:PRESet:CHANnel[:EXECute]
............................................................................ 179
INSTrument:CREate:DUPLicate
This command duplicates the currently selected measurement channel, i.e starts a
new measurement channel of the same type and with the identical measurement set-
tings. The name of the new channel is the same as the copied channel, extended by a
consecutive number (e.g. "Spectrum" -> "Spectrum 2").
The channel to be duplicated must be selected first using the
INST:SEL
command.
This command is not available if the MSRA/MSRT Master channel is selected.
Example:
INST:SEL 'Spectrum'
INST:CRE:DUPL
Duplicates the channel named 'Spectrum' and creates a new
measurement channel named 'Spectrum 2'.
Usage:
Event
INSTrument:CREate[:NEW]
<ChannelType>, <ChannelName>
This command adds an additional measurement channel. The number of measure-
ment channels you can configure at the same time depends on available memory.
Parameters:
<ChannelType>
Channel type of the new channel.
For a list of available channel types see
Activating Analog Demodulation Measurements