Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
255
User Manual 1173.9240.02 ─ 17
11.4.8 Configuring Demodulation
The following remote commands are required to configure the demodulation parame-
ters in a remote environment. The tasks for manual operation are described in
................................................................................ 255
.................................................................................256
Configuring the Demodulation Spectrum
.............................................................. 258
..................................................................................261
Defining the Scaling and Units
..............................................................................265
......................................................................................265
......................................................................................266
11.4.8.1
Basic Demodulation Settings
The basic demodulation measurement parameters define how the measurement is
performed.
Useful commands described elsewhere:
●
chapter 11.4.8.2, "Time Domain Zoom Settings"
Basic demodulation commands:
................................................................................ 255
.................................................................................... 255
[SENSe:]ADEMod:SQUelch[:STATe]
............................................................................... 256
...................................................................................256
[SENSe:]ADEMod<n>:AF:COUPling
<Coupling>
This command selects the coupling of the AF path of the analyzer in the specified win-
dow.
Parameters:
<Coupling>
AC | DC
*RST:
AC (PM); DC (FM)
Example:
ADEM:AF:COUP DC
Switches on DC coupling.
Manual operation:
See
[SENSe:]ADEMod:PM:RPOint[:X]
<Time>
This command determines the position where the phase of the PM-demodulated signal
is set to 0 rad. The maximum possible value depends on the measurement time
selected in the instrument; this value is output in response to the query
ADEM:PM:RPO:X? MAX
.
Configuring the Measurement