Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
206
User Manual 1173.9240.02 ─ 17
Parameters:
<SampleRate>
Range:
1 Hz to 10 GHz
*RST:
32 MHz
Example:
INP:DIQ:SRAT 200 MHz
Manual operation:
See
INPut:DIQ:SRATe:AUTO
<State>
If enabled, the sample rate of the digital I/Q input signal is set automatically by the con-
nected device.
This command is only available if the optional Digital Baseband Interface (R&S
FSW-
B17) is installed.
Parameters:
<State>
ON | OFF
*RST:
OFF
Manual operation:
See
11.4.2.5
Setting up Probes
Probes can be connected to the optional BASEBAND INPUT connectors, if the Analog
Baseband interface ( option R&S
FSW-B71) is installed.
[SENSe:]PROBe<p>:ID:PARTnumber?
............................................................................206
............................................................................... 207
.................................................................................207
............................................................................... 207
...............................................................................208
................................................................................ 208
[SENSe:]PROBe<p>:ID:PARTnumber?
Queries the R&S part number of the probe.
Suffix:
<p>
.
1 | 2 | 3
Selects the connector:
1 = Baseband Input I
2 = Baseband Input Q
3 = RF (currently not supported; use "1" with RF Input Connec-
tor setting "Baseband Input I")
Return values:
<PartNumber>
Part number in a string.
Usage:
Query only
Configuring the Measurement