Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
186
User Manual 1173.9240.02 ─ 17
Example:
MIX:LOP 16.0dBm
Manual operation:
See
[SENSe:]MIXer:SIGNal
<State>
This command specifies whether automatic signal detection is active or not.
Note that automatic signal identification is only available for measurements that per-
form frequency sweeps (not in vector signal analysis or the I/Q Analyzer, for instance).
Parameters:
<State>
OFF | ON | AUTO | ALL
OFF
No automatic signal detection is active.
ON
Automatic signal detection (Signal ID) is active.
AUTO
Automatic signal detection (Auto ID) is active.
ALL
Both automatic signal detection functions (Signal ID+Auto ID)
are active.
*RST:
OFF
Manual operation:
See
See
[SENSe:]MIXer:THReshold
<Value>
This command defines the maximum permissible level difference between test sweep
and reference sweep to be corrected during automatic comparison (see
Parameters:
<Value>
<numeric value>
Range:
0.1 dB to 100 dB
*RST:
10 dB
Example:
MIX:PORT 3
Manual operation:
See
Mixer Settings
The following commands are required to configure the band and specific mixer set-
tings.
[SENSe:]MIXer:FREQuency:HANDover
........................................................................... 187
[SENSe:]MIXer:FREQuency:STARt?
................................................................................187
................................................................................ 187
[SENSe:]MIXer:HARMonic:BAND:PRESet
........................................................................188
Configuring the Measurement