Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
252
User Manual 1173.9240.02 ─ 17
corresponds to the AM signal
PM
PM power signal
GP0 | GP1 | GP2 | GP3 | GP4 | GP5
For applications that process I/Q data, such as the I/Q Analyzer
or optional applications, and only if the Digital Baseband Inter-
face (R&S
FSW-B17) is available.
Defines triggering of the measurement directly via the LVDS
connector. The parameter specifies which general purpose bit (0
to 5) will provide the trigger data.
The assignment of the general purpose bits used by the Digital
IQ trigger to the LVDS connector pins is provided in
on page 97.
*RST:
IMMediate
Example:
TRIG:SOUR EXT
Selects the external trigger input as source of the trigger signal
Manual operation:
See
"Using the power sensor as an external trigger"
See
See
See
See
See
See
See
See
See
See
See
TRIGger[:SEQuence]:TIME:RINTerval
<Interval>
This command defines the repetition interval for the time trigger.
Parameters:
<Interval>
2.0 ms to 5000
Range:
2 ms to 5000 s
*RST:
1.0 s
Example:
TRIG:SOUR TIME
Selects the time trigger input for triggering.
TRIG:TIME:RINT 50
The sweep starts every 50 s.
Configuring the Measurement