![Rohde & Schwarz FSW Series User Manual Download Page 215](http://html.mh-extra.com/html/rohde-and-schwarz/fsw-series/fsw-series_user-manual_1477844215.webp)
Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
215
User Manual 1173.9240.02 ─ 17
Example:
PMET2:MTIM:AVER ON
Activates manual averaging.
PMET2:MTIM:AVER:COUN 8
Sets the number of readings to 8.
Manual operation:
See
"Average Count (Number of Readings)"
[SENSe:]PMETer<p>:MTIMe:AVERage[:STATe]
<State>
This command turns averaging for power sensor measurements on and off.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
PMET2:MTIM:AVER ON
Activates manual averaging.
Manual operation:
See
[SENSe:]PMETer<p>:ROFFset[:STATe]
<State>
This command includes or excludes the reference level offset of the analyzer for power
sensor measurements.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<State>
ON | 1
Includes the reference level offset in the results.
OFF | 0
Ignores the reference level offset.
*RST:
1
Example:
PMET2:ROFF OFF
Takes no offset into account for the measured power.
Manual operation:
See
[SENSe:]PMETer<p>[:STATe]
<State>
This command turns a power sensor on and off.
Suffix:
<p>
.
1...4
Power sensor index
Configuring the Measurement