Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
292
User Manual 1173.9240.02 ─ 17
Usage:
SCPI confirmed
11.7.2 Exporting Trace Results
Trace results can be exported to a file.
For more commands concerning data and results storage see the R&S
FSW User
Manual.
.........................................................................................292
........................................................................................ 292
..............................................................................................293
............................................................................................. 293
MMEMory:STORe<n>:TRACe
<Trace>, <FileName>
This command exports trace data from the specified window to an ASCII file.
Secure User Mode
In secure user mode, settings that are to be stored on the instrument are stored to vol-
atile memory, which is restricted to 256
MB. Thus, a "Memory full" error may occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details see "Protecting Data Using the Secure User Mode" in the "Data Manage-
ment" section of the R&S
FSW User Manual.
Parameters:
<Trace>
Number of the trace to be stored
(This parameter is ignored if the option "Export all Traces and all
Table Results" is activated in the Export configuration settings,
see
<FileName>
String containing the path and name of the target file.
Example:
MMEM:STOR1:TRAC 3,'C:\TEST.ASC'
Stores trace 3 from window 1 in the file
TEST.ASC
.
Usage:
SCPI confirmed
Manual operation:
See
FORMat:DEXPort:DSEParator
<Separator>
This command selects the decimal separator for data exported in ASCII format.
Retrieving Results