Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
253
User Manual 1173.9240.02 ─ 17
11.4.7.2
Configuring the Trigger Output
The following commands are required to send the trigger signal to one of the variable
TRIGGER INPUT/OUTPUT connectors. The tasks for manual operation are described
in
OUTPut:TRIGger<port>:DIRection
...................................................................................253
........................................................................................ 253
...................................................................................... 254
OUTPut:TRIGger<port>:PULSe:IMMediate
.......................................................................254
OUTPut:TRIGger<port>:PULSe:LENGth
.......................................................................... 254
OUTPut:TRIGger<port>:DIRection
<Direction>
This command selects the trigger direction.
Suffix:
<port>
.
Selects the trigger port to which the output is sent.
2 = trigger port 2 (front)
3 = trigger port 3 (rear)
Parameters:
<Direction>
INPut
Port works as an input.
OUTPut
Port works as an output.
*RST:
INPut
Manual operation:
See
OUTPut:TRIGger<port>:LEVel
<Level>
This command defines the level of the signal generated at the trigger output.
This command works only if you have selected a user defined output with
.
Suffix:
<port>
.
Selects the trigger port to which the output is sent.
2 = trigger port 2 (front)
3 = trigger port 3 (rear)
Parameters:
<Level>
HIGH
TTL signal.
LOW
0 V
*RST:
LOW
Manual operation:
See
See
Configuring the Measurement