Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
191
User Manual 1173.9240.02 ─ 17
[SENSe:]MIXer:PORTs
<PortType>
This command specifies whether the mixer is a 2-port or 3-port type.
Parameters:
<PortType>
2 | 3
*RST:
2
Example:
MIX:PORT 3
Manual operation:
See
[SENSe:]MIXer:RFOVerrange[:STATe]
<State>
If enabled, the band limits are extended beyond "RF Start" and "RF Stop" due to the
capabilities of the used harmonics.
Parameters:
<State>
ON | OFF
*RST:
OFF
Manual operation:
See
Conversion Loss Table Settings
The following settings are required to configure and manage conversion loss tables.
....................................................................................191
..................................................................................... 192
[SENSe:]CORRection:CVL:CATAlog?
..............................................................................192
................................................................................... 192
[SENSe:]CORRection:CVL:COMMent
..............................................................................193
.................................................................................... 193
[SENSe:]CORRection:CVL:HARMonic
............................................................................. 194
.................................................................................... 194
.................................................................................. 194
.................................................................................. 195
[SENSe:]CORRection:CVL:SNUMber
.............................................................................. 195
[SENSe:]CORRection:CVL:BAND
<Type>
This command defines the waveguide band for which the conversion loss table is to be
used. This setting is checked against the current mixer setting before the table can be
assigned to the range.
Before this command can be performed, the conversion loss table must be selected
(see
This command is only available with option B21 (External Mixer) installed.
Configuring the Measurement