9 ACP Measurement
Trace/Detector
other analyzer settings.
The detector choices are:
•
The Normal detector determines the peak of CW-like signals, and it yields alternating maximums
and minimums of noise-like signals. This is also referred to as Rosenfell detection.
•
The Average detector determines the average of the signal within the data range. The averaging
method is Power (RMS).
•
The Peak detector determines the maximum of the signal within the data range.
•
The Sample detector indicates the instantaneous level of the signal at the center of the data
represented by each display point.
•
The Negative Peak detector determines the minimum of the signal within the data range.
Because they may not find a spectral component's true peak, neither average nor sample detectors
measure amplitudes of CW signals as accurately as peak or normal, but they do measure noise
without the biases of peak detection.
When a detector selection is made, the menu returns to the previous menu.
Couplings
When Detector setting is “Auto” ([:SENSe]:ACPower:DETector:AUTO?), Detector is set to what the
Radio Standard defaults states (see detector section) for all conditions of Trace Type and for all
traces. When set to Manual, all Traces use the same detector type. When Average State = Off then
Trace Types AVERage, MaxHold and MinHold will not function, since Averaging is required to be ‘on’
for them to operate.
Only one detector type for all 3 traces is allowed.
When Meas Method is RBW or FAST, Detector is disabled.
Preset
AVERage
State Saved
Saved in instrument state.
Range
Normal|Average|Peak|Sample|Negative Peak
Backwards
Compatibility SCPI
[:SENSe]:ACPR:SWEep:DETector[:FUNCtion]
Initial S/W Revision
Prior to A.02.00
Modified at S/W Revision
A.02.00, A.03.00
W-CDMA & HSPA Measurement Applications Reference
965
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...