12 Code Domain Measurement
View/Display
Graticule
Pressing Graticule turns the display graticule On or Off. It also turns the graticule y-axis annotation on and
off.
Key Path
View/Display, Display
Remote Command
:DISPlay:WINDow[1]:TRACe:GRATicule:GRID[:STATe] OFF|ON|0|1
:DISPlay:WINDow[1]:TRACe:GRATicule:GRID[:STATe]?
Example
DISP:WIND:TRAC:GRAT:GRID OFF
Notes
The graticule is the set of horizontal and vertical lines that make up the grid/divisions for the x-axis
and y-axis.
Preset
On
State Saved
Saved in instrument state
Initial S/W Revision
Prior to A.02.00
System Display Settings
These settings are "Mode Global" – they affect all modes and measurements and are reset only by Restore
Misc Defaults or Restore System Defaults under System.
Key Path
View/Display, Display
Initial S/W Revision
Prior to A.02.00
Annotation Local Settings
This is a Mode Global override of the meas local annotation settings. When it is All Off, it forces
ScreenAnnotation, Meas Bar, Trace, and Active Function Values settings to be OFF for all measurements
in all modes. This provides the security based "annotation off" function of previous analyzers; hence it uses
the legacy SCPI command.
When it is All Off, the Screen, Meas Bar, Trace, and Active Function Values keys under the Display,
Annotation menu are grayed out and forced to Off. When Local Settings is selected, you are able to set the
local annotation settings on a measurement by measurement basis.
Key Path
View/Display, Display, System Display Settings
Remote Command
:DISPlay:WINDow[1]:ANNotation[:ALL] OFF|ON|0|1
:DISPlay:WINDow[1]:ANNotation[:ALL]?
Example
:DISP:WIND:ANN OFF
Preset
On (Set by Restore Misc Defaults)
State Saved
Not saved in instrument state.
Backwards Compatibility
Notes
The WINDow parameter and optional subopcode is included for backwards compatibility but ignored
– all windows are equally affected.
Initial S/W Revision
Prior to A.02.00
1952
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...