![Keysight E6640A Measurement Applications User & Programmer Reference Download Page 2011](http://html1.mh-extra.com/html/keysight/e6640a/e6640a_measurement-applications-user-and-programmer-reference_19616452011.webp)
13 Modulation Accuracy (Composite EVM) Measurement
AMPTD Y Scale
:DISPlay:RHO:VIEW6:WINDow3:TRACe:Y[:SCALe]:PDIVision?
Example
DISP:RHO:VIEW6:WIND3:TRAC:Y:PDIV 10.0
DISP:RHO:VIEW6:WIND3:TRAC:Y:PDIV?
Couplings
If Auto Scaling is set to On, this value is automatically determined by the measurement result. When
you set this value manually, Auto Scaling is automatically set to Off.
Preset
1.0
State Saved
Saved in instrument state.
Min
1 Hz
Max
1 GHz
Initial S/W Revision
Prior to A.02.00
Scale/Div: Code Domain Power View – Power Bar Graph Window
Sets the sensitivity result for the code domain power measurement in the code domain power graph
window.
Key Path
AMPTD Y Scale
Mode
WCDMA
Remote Command
:DISPlay:RHO:VIEW3:WINDow[1]:TRACe:Y[:SCALe]:PDIVision <real>
:DISPlay:RHO:VIEW3:WINDow[1]:TRACe:Y[:SCALe]:PDIVision?
Example
DISP:RHO:VIEW3:WIND:TRAC:Y:PDIV 10.0
DISP:RHO:VIEW3:WIND:TRAC:Y:PDIV?
Preset
10.00
State Saved
Saved in instrument state.
Min
0.10
Max
20.00
Initial S/W Revision
Prior to A.02.00
Ref Position
Positions the Y-axis scale reference level at the top, center or bottom of the display. Changing the
reference position does not change the reference level value.
Details of the Ref Position functionality depend on the selected view and window, as described in the
following sections:
•
"Ref Position: I/Q Error View - EVM Window" on page 2012
•
"Ref Position: I/Q Error View – Mag Error Window" on page 2012
•
"Ref Position: I/Q Error View – Phase Error Window" on page 2012
•
"Ref Position: Slot CDE/EVM View – Evm Window" on page 2013
W-CDMA & HSPA Measurement Applications Reference
2011
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...