![Keysight E6640A Measurement Applications User & Programmer Reference Download Page 186](http://html1.mh-extra.com/html/keysight/e6640a/e6640a_measurement-applications-user-and-programmer-reference_1961645186.webp)
3 Programming the Test Set
Common Commands
Service Request Enable
This command enables the desired bits of the service request enable register.
The query returns the value of the register, indicating which bits are currently enabled.
Remote Command
*SRE <integer>
*SRE?
Example
*SRE 22 Enables bits 1, 2, and 4 in the service request enable register.
Notes
For related commands, see the STATus subsystem and SYSTem:ERRor[:NEXT]? commands.
Preset
0
Min
0
Max
255
Status Bits/OPC
dependencies
Service Request Enable Register (all bits, 0 – 7).
Initial S/W Revision
Prior to A.02.00
Status Byte Query
Returns the value of the status byte register without erasing its contents.
Remote Command
*STB?
Example
*STB? Returns a decimal value for the bits in the status byte register.
For example, if a 16 is returned, it indicates that bit 5 is set and one of the conditions monitored in
the standard event status register is set.
Notes
See related command *CLS.
Status Bits/OPC
dependencies
Status Byte Register (all bits, 0 – 7).
Initial S/W Revision
Prior to A.02.00
Trigger
This command triggers the instrument. Use the :TRIGger[:SEQuence]:SOURce command to select the
trigger source.
Key Path
No equivalent key. See related keys Single and Restart.
Remote Command
*TRG
Example
*TRG Triggers the instrument to take a sweep or start a measurement, depending on the current
instrument settings.
Notes
See related command :INITiate:IMMediate.
Initial S/W Revision
Prior to A.02.00
186
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...