14 QPSK EVM Measurement
Meas
measurement and you send READ:ACP? it will initiate a new measurement using the same instrument settings.
•
Initiates the measurement and puts valid data into the output buffer. If a measurement other than the current one is
specified, the instrument will switch to that measurement before it initiates the measurement and returns results.
•
For example, suppose you have previously initiated the ACP measurement, but now you are running the channel
power measurement. Then you send READ:ACP? It will change from channel power back to ACP and, using the
previous ACP settings, will initiate the measurement and return results.
•
Blocks other SCPI communication, waiting until the measurement is complete before returning the results
•
If the optional [n] value is not included, or is set to 1, the scalar measurement results will be returned. If the [n] value
is set to a value other than 1, the selected trace data results will be returned. See each command for details of what
types of scalar results or trace data results are available. The binary data formats should be used when handling large
blocks of data since they are smaller and faster than the ASCII format. (FORMat:DATA)
Initial S/W Revision
Prior to A.02.00
Current Measurement Query (Remote Command Only)
This command returns the name of the measurement that is currently running.
Remote Command
:CONFigure?
Example
CONF?
Initial S/W Revision
Prior to A.02.00
Limit Test Current Results (Remote Command Only)
Queries the status of the current measurement limit testing. It returns a 0 if the measured results pass
when compared with the current limits. It returns a 1 if the measured results fail any limit tests.
Remote Command
:CALCulate:CLIMits:FAIL?
Example
CALC:CLIM:FAIL? queries the current measurement to see if it fails the defined limits.
Returns a 0 or 1: 0 it passes, 1 it fails.
Initial S/W Revision
Prior to A.02.00
Data Query (Remote Command Only)
Returns the designated measurement data for the currently selected measurement and subopcode.
n = any valid subopcode for the current measurement. See the measurement command results table for
your current measurement, for information about what data is returned for the subopcodes.
This command uses the data setting specified by the FORMat:BORDer and FORMat:DATA commands and
can return real or ASCII data. (See the format command descriptions under Input/Output in the Analyzer
Setup section.)
W-CDMA & HSPA Measurement Applications Reference
2389
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...