3 Programming the Test Set
STATus Subsystem
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Calibration Extended Needed Register
"Questionable Calibration Extended Needed Condition " on page 162
"Questionable Calibration Extended Needed Enable " on page 162
"Questionable Calibration Extended Needed Event Query " on page 163
"Questionable Calibration Extended Needed Negative Transition " on page 163
"Questionable Calibration Extended Needed Positive Transition " on page 164
Questionable Calibration Extended Needed Condition
This query returns the decimal value of the sum of the bits in the Questionable Calibration Extended
Needed Condition register.
The data in this register is continuously updated and reflects the current conditions.
Mode
All
Remote Command
:STATus:QUEStionable:CALibration:EXTended:NEEDed:CONDition?
Example
STAT:QUES:CAL:EXT:NEED:COND?
Preset
0
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Calibration Extended Needed Enable
This command determines which bits in the Questionable Calibration Extended Needed Condition Register
will set bits in the Questionable Calibration Extended Needed Event register, which also sets bit 14 of the
Questionable Calibration Register. The variable <integer> is the sum of the decimal values of the bits you
want to enable.
Mode
All
Remote Command
:STATus:QUEStionable:CALibration:EXTended:NEEDed:ENABle <integer>
:STATus:QUEStionable:CALibration:EXTended:NEEDed:ENABle?
Example
STAT:QUES:CAL:EXT:NEED:ENAB 2 Can be used to query if an EMI conducted alignment is needed.
Preset
32767
Min
0
162
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...