Trace/Detector
•
Normal–the detector determines the peak of the CW-like signals, and it yields alternating maximums
and minimums of noise-like signals. This is also referred to as Rosenfell detection.
•
Average–the detector determines the average of the signal within the sweep points. The averaging
method depends upon the Average Type selection (voltage, power or log scales).
•
Peak–the detector determines the maximum of the signal within the sweep points.
•
Sample–the detector indicates the instantaneous level of the signal at the center of the sweep points
represented by each display point.
•
Negative Peak–the detector determines the minimum of the signal within the sweep points.
Key Path
Detector
Initial S/W Revision
Prior to A.02.00
Auto
Sets the detector for the currently selected trace to Auto.
Key Path
Trace/Detector
Mode
SA, WCDMA, C2K, WIMAX OFDMA, 1xEVDO, DVB-T/H, DTMB (CTTB), ISDB-T, CMMB, LTE,
LTETDD, Digital Cable TV, WLAN, MSR, LTEAFDD, LTEATDD
Remote Command
[:SENSe]:CHPower:DETector:AUTO ON|OFF|1|0
[:SENSe]:CHPower:DETector:AUTO?
Example
CHP:DET:AUTO ON
CHP:DET:AUTO?
Couplings
When Detector setting is “Auto” ([:SENSe]:CHPower:DETector:AUTO?), Detector
([:SENSe]:CHPower:DETector[:FUNCtion]?) switches aligning with the switch of this parameter:
“NORMal” with Clear Write, “AVERage” with AVERage, “POSitive (peak)” with MAXHold, and
“NEGative (peak)” with MINHold.
Preset
Others: ON
DVB-T/H, DTMB (CTTB), ISDB-T, CMMB, Digital Cable TV: OFF
State Saved
Saved in instrument state.
Initial S/W Revision
Prior to A.02.00
Modified at S/W Revision
A.02.00, A.03.00
Detector Selection
Selects a detector to be used by the analyzer for the current measurement.
Key Path
Trace/Detector
Mode
SA, WCDMA, C2K, WIMAX OFDMA, 1xEVDO, DVB-T/H, DTMB (CTTB), ISDB-T, CMMB, LTE,
LTETDD, Digital Cable TV, WLAN, MSR, LTEAFDD, LTEATDD
W-CDMA & HSPA Measurement Applications Reference
631
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...