![Keysight E6640A Measurement Applications User & Programmer Reference Download Page 218](http://html1.mh-extra.com/html/keysight/e6640a/e6640a_measurement-applications-user-and-programmer-reference_1961645218.webp)
4 Input/Output Functions
Input/Output
Notes
For PSA compatibility the command form is provided and is directly mapped to
[:SENSe]:ROSCillator:SOURce:TYPE
Note: The SCPI command does nothing for non-controlling instances.
Initial S/W Revision
Prior to A.02.00
Sense
If Sense is selected, the instrument checks whether a signal is present at the external reference
connector. If it senses a signal within 5 ppm of the External Ref Freq (as set on the External Ref Freq
softkey), it will use this signal as an External Reference. If it senses a 1 pulse per second signal, it will use
this signal to adjust the internal reference by adjusting the User setting of the Timebase DAC. When no
signal is present, it automatically switches to the internal reference.
Key Path
Input/Output, Freq Ref In
Example
:ROSC:SOUR:TYPE SENS
Couplings
If set to SENSe and the analyzer senses a 1 pulse per second signal, it sets the System, Alignments,
Timebase DAC setting to “User”. This setting survives Preset and Power Cycle but is set to
“Calibrated” on a System, Restore Defaults, Align or a System, Restore Defaults, All
Readback
Sense
Initial S/W Revision
Prior to A.02.00
Modified at S/W Revision
A.14.00
Internal
The internal reference is used. A 1 pps signal at the EXT REF IN port, or a signal there between 1 and 50
MHz, will cause a warning triangle to appear in the settings panel next to the word “INTERNAL”, but will
otherwise be ignored.
Key Path
Input/Output, Freq Ref In
Example
:ROSC:SOUR:TYPE INT
Readback
Internal
Initial S/W Revision
Prior to A.02.00
Modified at S/W Revision
A.14.00
External
The external reference is used.
Key Path
Input/Output, Freq Ref In
Example
:ROSC:SOUR:TYPE EXT
Readback
External
Initial S/W Revision
Prior to A.02.00
218
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...