7 Trigger Functions
Trigger
Notes
and RFB triggers.
The legacy :TRIGger[:SEQuence]:OFFSet command is supported for the VIDeo, LINE, EXT1, and EXT2
triggers.
Initial S/W Revision
Prior to A.02.00
Zero Span Delay Comp On/Off
In zero span, there is a natural delay in the signal path, which comes from the RBW filter. This is usually
desirable, as it allows you to trigger on events and also see those events, because the signal is delayed
from the trigger event. However, in some cases it is desirable to eliminate this delay, so that trigger
events line up exactly with the zero time point in zero span. You can use the Zero Span Delay Comp On/Off
feature to enable or disable zero span delay compensation.
Key Path
Trigger, External 1
Remote Command
:TRIGger[:SEQuence]:EXTernal1:DELay:COMPensation OFF|ON|0|1
:TRIGger[:SEQuence]:EXTernal1:DELay:COMPensation?
Example
TRIG:EXT1:DEL:COMP ON
Dependencies
No effect except in zero-span, but not locked out in nonzero spans.
Blanked in modes that do not support zero-span measurements. If the SCPI command is sent when
the key is blanked, an error is returned: –221, “Settings conflict; Feature not supported for this
measurement”
In analyzers shipping N9060A, this feature requires N9060A–7FP.
Preset
OFF
State Saved
Saved in instrument state
Initial S/W Revision
A.11.00
External 2
Pressing this key, when it is not selected, selects an external input signal as the trigger. A new
sweep/measurement will start when the external trigger condition is met using the external 2 input
connector. The external trigger 2 input connector is on the rear panel.
Pressing this key, when it is already selected, accesses the external 2 trigger setup menu.
Key Path
Trigger
Example
TRIG:SOUR EXT2 Swept SA measurement
TRIG:<meas>:SOUR EXT2 Measurements other than Swept SA
Dependencies
In some models, there is no second External input. In these models, the External 2 key is blanked and
the EXTernal2 parameter will generate a “Hardware missing; Not available for this model number”
message.
Grayed out if in use by Point Trigger in the Source Setup menu.
Forced to Free Run if already selected and Point Trigger is set to External 2.
340
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...