![Keysight E6640A Measurement Applications User & Programmer Reference Download Page 714](http://html1.mh-extra.com/html/keysight/e6640a/e6640a_measurement-applications-user-and-programmer-reference_1961645714.webp)
9 ACP Measurement
Meas
ASCii - Amplitude values are in ASCII, in the current Y Axis Unit, one ASCII character per digit, values
separated by commas, each value in the form:
SX.YYYYYEsZZ
Where:
S = sign (+ or -)
X = one digit to left of decimal point
Y = 5 digits to right of decimal point
E = E, exponent header
s = sign of exponent (+ or -)
ZZ = two digit exponent
REAL,32 - Binary 32-bit real values in the current Y Axis Unit, in a definite length block.
REAL,64 - Binary 64-bit real values in the current Y Axis Unit, in a definite length block.
Format Data: Byte Order (Remote Command Only)
This command selects the binary data byte order for data transfer and other queries. It controls whether
binary data is transferred in normal or swapped mode. This command affects only the byte order for setting
and querying trace data for the :TRACe[:DATA], TRACe[:DATA]? , :CALCulate:DATA[n]? and
FETCh:SANalyzer[n]? commands and queries.
By definition any command that says it uses FORMat:DATA uses any format supported by FORMat:DATA.
The NORMal order is a byte sequence that begins with the most significant byte (MSB) first, and ends with
the least significant byte (LSB) last in the sequence: 1|2|3|4. SWAPped order is when the byte sequence
begins with the LSB first, and ends with the MSB last in the sequence: 4|3|2|1.
Remote Command
:FORMat:BORDer NORMal|SWAPped
:FORMat:BORDer?
Preset
NORMal
Initial S/W Revision
Prior to A.02.00
714
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...