11 Occupied Bandwidth Measurement
Sweep/Control
menu is grayed out.
Preset
Off
LTETDD: On
State Saved
Saved in instrument state
Range
On|Off
Backwards
Compatibility SCPI
[:SENSe]:SWEep:TIME:GATE[:STATe] ESA compatibility
Backwards Compatibility
Notes
In ESA, Trig Delay (On) and Gate (On) could not be active at the same time.. This dependency does
not exist in PSA or in the X-Series.
Initial S/W Revision
Prior to A.02.00
Gate View On/Off
Turning on Gate View in the Swept SA measurement provides a single-window gate view display..
Turning on Gate View in other measurements shows the split-screen Gate View. In these measurements,
when the Gate View is on, the regular view of the current measurement traces and results are reduced
vertically to about 70% of the regular height. The Zero Span window, showing the positions of the Gate, is
shown between the Measurement Bar and the reduced measurement window. By reducing the height of
the measurement window, some of the annotation on the Data Display may not fit and is not shown.
Key Path
Sweep/Control, Gate
Remote Command
[:SENSe]:SWEep:EGATe:VIEW ON|OFF|1|0
[:SENSe]:SWEep:EGATe:VIEW?
Example
SWE:EGAT:VIEW ON turns on the gate view.
Dependencies
In the Swept SA measurement:
In Gate View, the regular Acq Time key is grayed out . When pressed, the grayed out key puts up the
informational message "Use Gate View Sweep Time in the Gate menu."
In the other measurements:
When you turn Gate View on, the lower window takes on the current state of the instrument. Upon
leaving Gate View, the instrument takes on the state of the lower window.
When you turn Gate View on, the upper window Acquisition Time is set to the gate view acquisition
time.
Couplings
These couplings apply to the Swept SA measurement:
•
When Gate View is turned on, the instrument is set to Zero Span.
•
Gate View automatically turns off whenever a Span other than Zero is selected.
•
Gate View automatically turns off if you press the Last Span key while in Gate View, and the
instrument returns to the Span it was in before entering Gate View (even if that is Zero Span).
•
When Gate View is turned on, the sweep time used is the gate view sweep time. This is set
according to the rules in section
"Gate View Setup " on page 1571
•
When Gate View is turned off, Sweep Time is set to the normal Swept SA measurement sweep
time.
1568
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...