![Keysight E6640A Measurement Applications User & Programmer Reference Download Page 1652](http://html1.mh-extra.com/html/keysight/e6640a/e6640a_measurement-applications-user-and-programmer-reference_19616451652.webp)
12 Code Domain Measurement
AMPTD Y Scale
0.125 V Peak
Set the channel gain state to 0.125 Volt Peak.
Key Path
AMPTD Y Scale, I Range | Q Range
Initial S/W Revision
Prior to A.02.00
Q Range Value
This is the internal gain range for the Q channel. See I/Q Gain Ranges. The Q Range only applies to Input
Path Q Only and Ind I/Q. For input I+jQ the I Range determines both I and Q channel range settings.
Key Path
AMPTD Y Scale, Range
Remote Command
[:SENSe]:VOLTage:IQ:Q:RANGe[:UPPer] <voltage>
[:SENSe]:VOLTage:IQ:Q:RANGe[:UPPer]?
Example
Set the Q Range to 0.5 V Peak
VOLT:IQ:Q:RANG 0.5 V
Notes
The numeric entries are mapped to the smallest gain range whose break point is greater than or
equal to the value, or 1 V Peak if the value is greater than 1 V.
The Q Range is only used for Input Path Q Only and Ind I/Q. For input I+jQ the I Range determines
both I and Q channel range settings.
Couplings
When Q Same as I is On, the I Range value will be copied to the Q Range and the range value keys
are disabled.
Changing the value will also set Range = Man.
Preset
1 V Peak
State Saved
Saved in instrument state
Range
1 V Peak | 0.5 V Peak | 0.25 V Peak | 0.125 V Peak
Min
0.125 V
Max
1 V
Initial S/W Revision
Prior to A.02.00
Remote Command
[:SENSe]:POWer:IQ:Q:RANGe[:UPPer] <ampl>
[:SENSe]:POWer:IQ:Q:RANGe[:UPPer]?
Example
Will set the Q Range to 0.5 V Peak when Reference Z is 50
Ω
, and to 1.0 V Peak when Reference Z
is 75
Ω
.
POW:IQ:Q:RANG 4 dBm
Notes
The POWer form of the command is provided for convenience. It maps to the same underlying gain
range parameter as the VOLTage form of the command.
The Reference Z (not the Q channel Input Z) is used to convert the power to peak voltage, which is
then used to set the Q Range as with the VOLTage form of the command. The power values of the 4
range states (1V Peak, 0.5V Peak, 0.25V Peak, and 0.125V Peak) will vary with Reference Z. Here
1652
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...