10 Spectrum Emission Mask Measurement
Meas Setup
If the current mode is DVB-T/H, this value will be modified automatically according to the output
power of the transmitter which is less or more than 25W.
If the current mode is ISDB-T, this value will be modified automatically according to the limit type.
Preset
Automatically calculated
Modes (except MSR, LTEAFDD, LTEATDD and WLAN) without option N9060A–7FP: ON, ON, ON,
ON, ON, ON
Modes (except MSR, LTEAFDD, LTEATDD and WLAN) with option N9060A–7FP: ON, ON, ON, ON,
ON, ON, ON, ON, ON, ON, ON, ON
WLAN: ON, ON, ON, ON, ON, ON, ON, ON, ON, ON, ON, ON
State Saved
Saved in instrument state.
Min
TBD
Max
4000 s
Backwards
Compatibility SCPI
[:SENSe]:SEMask:OFFSet[1]|2:LIST:SWEep[:TIME]
Initial S/W Revision
Prior to A.02.00
Modified at S/W Revision
A.02.00, A.03.00, A.10.00, A.11.00, A.14.00
Offset Side
Specifies which offset side to measure.
You can turn off (not use) specific offsets with [:SENSe]:SEMask:OFFSet[n][:OUTer]:LIST:STATe.
•
BOTH – Both of the negative (lower) and positive (upper) sidebands
•
NEGative – Negative (lower) sideband only
•
POSitive – Positive (upper) sideband only
Missing values are not permitted; that is, if you want to change values 2 and 6, you must send all values up
to 6. Subsequent values will remain as they were. The query for this parameter always returns 12 values.
The figure below shows the relation between the negative/positive offset side setups and the upper/lower
results in the MSR and LTE-Advanced FDD/TDD.
1070
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...