15 Power Control Measurement
Meas
DC Coupled
Example
CALC:FPOW:POW1:DEF "DCCoupled=True"
Notes
The DC coupled parameter allows the user to specify whether the DC blocking capacitor is utilized. Set
parameter to true when measuring frequencies below 10 MHz.
Preset
False
Range
True (DC Coupled) or False (AC Coupled)
Default Unit
Boolean
Initial S/W
Revision
A.14.00
DetectorType
Example
CALC:FPOW:POW1:DEF "DetectorType=Peak"
Notes
Option FP2 is required.
The detector type parameter allows the user to choose whether a RMS average or peak value is used
during the measurement.
Preset
RmsAverage
Range
RmsAverage, Peak
Initial S/W
Revision
A.14.00
Do Noise Correction
Example
CALC:FPOW:POW1:DEF "DoNoiseCorrection=True"
Notes
When noise correction is enabled, the linear noise power contributed by the analyzer is subtracted from all
measurements. This effectively lowers the noise floor of the analyzer.
When noise correction is enabled, the first measurement for a given set of input parameters will take extra
time. This is because the analyzer takes an extra acquisition with the RF input disconnected from the
analyzer’s front end to measure the noise of just the analyzer. The measured noise floor is stored in a cache
so the noise acquisition will occur only once for the same state settings. In other words, if noise correction
was turned on and the analyzer made an acquisition at frequency A, then frequency B, and back again to
frequency A, the hidden initial noise floor acquisition would only occur for the first acquisition at frequency A
and the cached noise floor would be used the second time frequency A was measured.
Preset
False
Range
True (enable noise correction) or False (disable noise correction)
Default Unit
Boolean
Initial S/W
Revision
A.14.00
2654
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...