12 Code Domain Measurement
Source
:SOUR:FREQ:REF:STATe ON
Dependencies
This setting is unavailable, and is grayed out when the List Sequencer is turned ON.
Couplings
The frequency reference state is coupled to the frequency reference set immediate action. When the
reference set immediate action key is pressed, or the SCPI command issued, it turns the frequency
reference state ON.
Preset
0.00 Hz
OFF
Min
0.00 Hz
Max
Hardware Dependant:
Option 503 = 3.6 GHz
Option 504 = 3.8 GHz
Option 506 = 6.00 GHz
Initial S/W Revision
A.05.00
Freq Offset
Allows you to specify the frequency offset value. When the frequency offset state is ON, an annunciator is
displayed on the main source view to indicate this state to the user.
When the frequency offset is set to zero (0) and you set a new offset value, the displayed frequency value
will change as follows and the RF output frequency will not change:
Displayed value = output fre offset value
Where:
output frequency equals the original frequency entered under Source>Frequency>Frequency
offset value equals the value entered under Source>Frequency>Freq Offset
When the frequency offset is set to a value other than zero (0) and you enter a new frequency value under
Source>Frequency>Frequency, the displayed frequency will be the same as the value entered and the RF
output frequency will be equal to the value entered minus the offset value as follows:
Output frequency = entered frequency – offset frequency
Displayed frequency = output fre offset frequency
Displayed frequency = entered frequency
Where:
entered frequency equals the frequency entered under Source>Frequency>Frequency
offset frequency equals the value previously entered and set under Source>Frequency>Freq Offset
Key Path
Source, Frequency
Remote Command
:SOURce:FREQuency:OFFSet <freq>
W-CDMA & HSPA Measurement Applications Reference
1843
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...