![Keysight E6640A Measurement Applications User & Programmer Reference Download Page 1702](http://html1.mh-extra.com/html/keysight/e6640a/e6640a_measurement-applications-user-and-programmer-reference_19616451702.webp)
12 Code Domain Measurement
Meas
Do Spur Suppression
Example
CALC:FPOW:POW1:DEF "DoSpurSuppression=True"
Notes
When measuring very low level signals, or when large out-of-band inputs are input into the analyzer,
sometimes unwanted spurs and residuals can appear in the measured spectrum. Spur suppression is a
method to help minimize the levels of these internally generated spurs and residuals.
When spur suppression is enabled, the analyzer will automatically take two acquisitions using two different
internal analog LO frequencies. The FFT spectrums from both acquisitions are combined by taking the
minimum power between both traces on a per FFT bin basis. External signals will have the same amplitude
for both traces and therefore will return the expected amplitudes. However, low level spurs and residuals
generated internally to the analyzer tend to move to different FFT bins depending on the internal analog LO
frequency used, and therefore tend to be suppressed using this spur suppression method.
Because two acquisitions, rather than a single acquisition, are made when spur suppression is enabled, the
measurement time will always be slower when spur suppression is enabled.
Preset
False
Range
True (enable spur suppression) or False (disable spur suppression)
Default Unit
Boolean
Initial S/W
Revision
A.14.00
Electronic Attenuator Bypass
Example
CALC:FPOW:POW1:DEF "ElecAttBypass =False"
Notes
The electronic attenuation bypass parameter allows the user to either utilize or bypass the electronic
attenuator. The electronic attenuator is only available for frequencies up to 3.6 GHz. Set parameter to true
when using frequencies above 3.6 GHz and set the parameter to false when using the preamp.
Preset
True
Range
True (bypass electronic attenuator) or False (use electronic attenuator)
Default Unit
Boolean
Initial S/W
Revision
A.14.00
Electronic Attenuation
Example
CALC:FPOW:POW1:DEF "ElecAttenuation=10"
Notes
Option EA3 is required.
The electronic attenuation value parameter sets the amount of electrical attenuation from 0 to 24 dB (1
dB steps).
Set "ElecAttBypass=False" to make sure the electronic attenuator path is enabled.
Preset
0 dB
Range
0 – 24 dB (1 dB steps)
1702
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...