![Keysight E6640A Measurement Applications User & Programmer Reference Download Page 1882](http://html1.mh-extra.com/html/keysight/e6640a/e6640a_measurement-applications-user-and-programmer-reference_19616451882.webp)
12 Code Domain Measurement
Source
Marker 3
Sets marker 3 to be used for the pulse/RF blanking function.
Key Path
Source, Modulation Setup, ARB, Marker Utilities, Marker Routing, Pulse/RF Blank
Example
:SOUR:RAD:ARB:MDES:PULS M3
Initial S/W Revision
A.05.00
Marker 4
Sets marker 4 to be used for the pulse/RF blanking function.
Key Path
Source, Modulation Setup, ARB, Marker Utilities, Marker Routing, Pulse/RF Blank
Example
:SOUR:RAD:ARB:MDES:PULS M4
Initial S/W Revision
A.05.00
ALC Hold
Allows you to specify which marker is routed for use within the ALC hold function. The ALC hold marker
function holds the ALC circuitry at the average value of the sample points set by the marker.
The ALC hold function operates during the low periods of the marker signal. The marker polarity
determines when the marker signal is high. For positive polarity, this is during the marker points. For a
negative polarity, this is when there are no maker points.
Key Path
Source, Modulation Setup, ARB, Marker Utilities, Marker Routing
Remote Command
:SOURce:RADio:ARB:MDEStination:ALCHold NONE|M1|M2|M3|M4
:SOURce:RADio:ARB:MDEStination:ALCHold?
Example
:SOUR:RAD:ARB:MDES:ALCH NONE
Dependencies
When a new waveform is selected for playback the settings contained within the associated
waveform header file are applied to the ARB. The ALC hold setting is one of the values stored within
the header file. If the newly selected waveform file has an associated header file, the ALC hold setting
is updated with the value from the header file. The ALC hold setting will remain unchanged if the
newly selected waveform does not have an associated header file.
Range
None | M1 | M2 | M3 | M4
Initial S/W Revision
A.05.00
None
Sets no marker to be used for the ALC hold function, essentially turning the ALC hold function off.
Key Path
Source, Modulation Setup, ARB, Marker Utilities, Marker Routing, ALC Hold
Example
:SOUR:RAD:ARB:MDES:PULS NONE
Initial S/W Revision
A.05.00
1882
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...