13 Modulation Accuracy (Composite EVM) Measurement
Source
Key Path
Source, Modulation Setup, ARB, Waveform Utilities, Multi-Pack Licenses
Remote Command
:SYSTem:LKEY:WAVeform:CLEar <int>
or
:SYSTem:LICense[:FPACk]:WAVeform:CLEar <int>
Example
SYST:LKEY:WAV:CLE 1
or
:SYST:LIC:WAV:CLE 1
Notes
The second SCPI :SYSTem:LICense[:FPACk]:WAVeform:CLEar is provided to be consistent with the
style of Agilent signal sources. You can use either one of them.
Waveform slot number <int> is positive. If you attmpt to input a slot number less than or equals 0, an
error is generated.
If current module is not “TRX1” module, the key will grey out, and error message is generated “–221
Setting conflict; Not allowed on current module. Goto “TRX1” to operate multi-pack license” when
invoking SCPI.
Dependencies
This key is only available if the currently selected slot is in the trial state.
Initial S/W Revision
A.05.00
Lock Waveform in Slot
If the selected slot is in the trial state or the lock required state, the waveform that occupies the slot is
locked and permanently licensed.
Key Path
Source, Modulation Setup, ARB, Waveform Utilities, Multi-Pack Licenses
Remote Command
:SYSTem:LKEY:WAVeform:LOCK <int>
or
:SYSTem:LICense[:FPACk]:WAVeform:LOCK <int>
Example
SYST:LKEY:WAV:LOCK 1
or
SYST:LIC:WAV:LOCK 1
Notes
The second SCPI :SYSTem:LICense[:FPACk]:WAVeform:LOCK is provided to be consistent with the
style of Agilent signal sources. You can use either one of them.
Waveform slot number <int> is positive. If you attmpt to input a slot number less than or equals 0, an
error is generated.
If current module is not “TRX1” module, the key will grey out, and error message is generated “–221
Setting conflict; Not allowed on current module. Goto “TRX1” to operate multi-pack license” when
invoking SCPI.
Dependencies
This key is only available if the currently selected slot is in the trial state or the lock required state.
Initial S/W Revision
A.05.00
W-CDMA & HSPA Measurement Applications Reference
2233
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...