10 Spectrum Emission Mask Measurement
Sweep/Control
Dependencies
In the Swept SA measurement:
In Gate View, the regular Acq Time key is grayed out . When pressed, the grayed out key puts up the
informational message "Use Gate View Sweep Time in the Gate menu."
In the other measurements:
When you turn Gate View on, the lower window takes on the current state of the instrument. Upon
leaving Gate View, the instrument takes on the state of the lower window.
When you turn Gate View on, the upper window Acquisition Time is set to the gate view acquisition
time.
Couplings
These couplings apply to the Swept SA measurement:
•
When Gate View is turned on, the instrument is set to Zero Span.
•
Gate View automatically turns off whenever a Span other than Zero is selected.
•
Gate View automatically turns off if you press the Last Span key while in Gate View, and the
instrument returns to the Span it was in before entering Gate View (even if that is Zero Span).
•
When Gate View is turned on, the sweep time used is the gate view sweep time. This is set
according to the rules in section
"Gate View Setup " on page 1571
•
When Gate View is turned off, Sweep Time is set to the normal Swept SA measurement sweep
time.
•
If Gate View is on and Gate is off, then turning on Gate turns off Gate View.
Preset
OFF
State Saved
Saved in instrument state
Range
On|Off
Initial S/W Revision
Prior to A.02.00
A sample of the Gate View screen in the Swept SA measurement is shown in the following graphic :
1266
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...