13 Modulation Accuracy (Composite EVM) Measurement
Source
This command writes a waveform sequence file to the hard disk. You must specify the waveform sequence
file path and filename which will be saved on the hard disk, and the waveform segment file path and name
which will be nested into the waveform sequence file. You can utilize mass storage unit specifier (MSUS)
“NVWFM” or use a real full path representation. See the example below. MSUS “NVWFM” is mapped to D:
VARB directory on test set hard disk.
Any number of segments, up to a segment count limit of 64, can be used to create a sequence. Repeated
segments are included in the count limit.
Each waveform segment name string length upper limit is 128 chars. Please do NOT insert waveform which
name string exceeds 128 chars.
The internal source does not support nesting one waveform sequence file into another waveform
sequence file.
Remote Command
:SOURce:RADio:ARB:SEQuence[:MWAVeform] <filename>, <waveform1>, <reps>,
NONE | M1 | M2 | M3 | M4 | M1M2 | M1M3 | M1M4 | M2M3 | M2M4 | M3M4 |
M1M2M3 | M1M2M4 | M1M3M4 | M2M3M4 | M1M2M3M4 | ALL, {<waveform2>,
<reps>, NONE | M1 | M2 | M3 | M4 | M1M2 | M1M3 | M1M4 | M2M3 | M2M4 |
M3M4 | M1M2M3 | M1M2M4 | M1M3M4 | M2M3M4 | M1M2M3M4 | ALL, } …
(For additional description of each item, see Notes below
“For Setup
SCPI”.)
:SOURce:RADio:ARB:SEQuence[:MWAVeform]? <filename>
(For additional description of each item, see Notes
below.)
Example
For setup:
>:SOUR:RAD:ARB:SEQ “NVWFM:testSeq1.seq”, “NVWFM:wfmSegment1.wfm”, 10, M2M3M4,
“NVWFM:wfmSegment2.wfm”, 20, M1M3
Or
>:SOUR:RAD:ARB:SEQ “D: VARB\testSeq1.seq”, “ D: VARB\wfmSegment1.wfm”, 10, M2M3M4, “ D:
VARB\wfmSegment2.wfm”, 20, M1M3
For query, must specify which waveform sequence file to query.
>:SOUR:RAD:ARB:SEQ? “NVWFM:testSeq1.seq”
Or
>:SOUR:RAD:ARB:SEQ? “D: VARB\testSeq1.seq”,
Notes
For Setup SCPI
For the Setup SCPI command, the parameters are:
<filename> - String Type
This variable specifies the path and name for the waveform sequence file. The path supports MSUS
(NVWFM) or a real full path representation. See example.
<waveform1> - String Type
This variable specifies the path and name of the first existing waveform segment. The path supports
MSUS (NVWFM) or a real full path representation. See example.
The segment file must reside within ARB playback memory before it can be played by the ARB player.
<reps> - Integer Type
2224
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...