14 QPSK EVM Measurement
Source
when the ARB memory is full, the copy ceases, and an error is generated.
If you specify a directory over SCPI, but the directory does not exist, an error is generated.
If the ARB is ON, a user then loads or deletes file to ARB memory, the playing waveform segment
may not keep phase continuity during the ARB memory operation. The waveform will be replayed
after the ARB operation is finished.
Initial S/W Revision
A.05.00
Change Directory...
Allows you to change the currently selected directory on the hard disk. Pressing this key opens a standard
windows change directory dialog allowing you to select the new directory of interest.
The current directory is used for manually loading waveform segments into ARB memory for playback, and
as a search location for waveform segments that are required to be loaded into ARB memory for playback
of a waveform sequence or a list sequence.
Key Path
Source, Modulation Setup, ARB, Select Waveform, Segments on Hard Disk
Notes
No remote command, SCPIfront panel only.
Initial S/W Revision
A.05.00
Default Directory...
Allows you to change the default directory. It is used as a search location for waveform segments that are
required to be loaded into ARB memory for playback of a waveform sequence, and as a search location for
selecting waveforms using SCPI.
Key Path
Source, Modulation Setup, ARB, Select Waveform, Segments on Hard Disk
Remote Command
:SOURce:RADio:ARB:DEFault:DIRectory <string>
:SOURce:RADio:ARB: DEFault:DIRectory?
Example
:SOUR:RAD:ARB:DEF:DIR “D:\ArbFiles”
:SOUR:RAD:ARB:DEF:DIR?
State Saved
Persistent, survives a power cycle and a preset but not saved in the instrument state
Initial S/W Revision
A.05.00
Segments in ARB Memory
Allows you access to the sub-menus for managing the files within ARB memory.
Key Path
Source, Modulation Setup, ARB, Select Waveform
Initial S/W Revision
A.05.00
W-CDMA & HSPA Measurement Applications Reference
2505
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...