![Keysight E6640A Measurement Applications User & Programmer Reference Download Page 2533](http://html1.mh-extra.com/html/keysight/e6640a/e6640a_measurement-applications-user-and-programmer-reference_19616452533.webp)
14 QPSK EVM Measurement
Source
Slot Status Query (Remote Command Only)
Returns the status of the specified slot.
Remote Command
:SYSTem:LKEY:WAVeform:STATus? <int>
or
:SYSTem:LICense[:FPACk]:WAVeform:STATus? <int>
Example
:SYST:LKEY:WAV:STAT? 1
<"Locked"
or
:SYST:LIC:WAV:STAT? 1
<"Locked"
Notes
The second SCPI :SYSTem:LICense[:FPACk]:WAVeform:STATus is provided to be consistent with the
style of Agilent signal sources. You can use either one of them.
Waveform slot number <int> is positive. If you attmpt to input a slot number less than or equals 0, an
error is generated.
Result type is string. If input slot number exceeds total available slot number, “Nonexistent” is
returned.
Range
"Locked" | "Available" | “Trail” | “LockRequired” | “Nonexistent”
Initial S/W Revision
A.05.00
Slots Free Query (Remote Command Only)
Returns the number of license slots free.
Remote Command
:SYSTem:LKEY:WAVeform:FREE?
or
:SYSTem:LICense[:FPACk]:WAVeform:FREE?
Example
:SYST:LKEY:WAV:FREE?
or
:SYST:LIC:WAV:FREE?
Notes
The second SCPI :SYSTem:LICense[:FPACk]:WAVeform:FREE is provided to be consistent with the
style of Agilent signal sources. You can use either one of them.
Initial S/W Revision
A.05.00
Slot Used Query (Remote Command Only)
Returns the number of license slots used.
Remote Command
:SYSTem:LKEY:WAVeform:USED?
or
:SYSTem:LICense[:FPACk]:WAVeform:USED?
Example
:SYST:LKEY:WAV:USED?
W-CDMA & HSPA Measurement Applications Reference
2533
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...