![Keysight E6640A Measurement Applications User & Programmer Reference Download Page 610](http://html1.mh-extra.com/html/keysight/e6640a/e6640a_measurement-applications-user-and-programmer-reference_1961645610.webp)
8 Channel Power Measurement
Sweep/Control
events line up exactly with the zero time point in zero span. You can use the Zero Span Delay Comp On/Off
feature to enable or disable zero span delay compensation.
Key Path
Trigger, External 2
Remote Command
:TRIGger[:SEQuence]:EXTernal2:DELay:COMPensation OFF|ON|0|1
:TRIGger[:SEQuence]:EXTernal2:DELay:COMPensation?
Example
TRIG:EXT2:DEL:COMP ON
Dependencies
No effect except in zero-span, but not locked out in nonzero spans.
Blanked in modes that do not support zero-span measurements. If the SCPI command is sent when
the key is blanked, an error is returned: –221, “Settings conflict; Feature not supported for this
measurement”
In analyzers shipping N9060A, this feature requires N9060A–7FP.
Preset
OFF
State Saved
Saved in instrument state
Initial S/W Revision
A.11.00
RF Burst
Pressing this key, when it is not selected, selects the RF Burst as the trigger. A new sweep/measurement
will start when an RF burst envelope signal is identified from the signal at the RF Input connector. Pressing
this key, when it is already selected, accesses the RF Burst trigger setup menu.
In some models, a variety of burst trigger circuitry is available, resulting in various available burst trigger
bandwidths. The analyzer automatically chooses the appropriate trigger path based on the hardware
configuration and other settings of the analyzer.
Key Path
Trigger
Example
TRIG:SOUR RFB Swept SA measurement
TRIG:<meas>:SOUR RFB Measurements other than Swept SA
State Saved
Saved in instrument state
Status Bits/OPC
dependencies
The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping or
Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger event
occurs and all the applicable trigger criteria have been met). A corresponding pop-up message
("Waiting for trigger") is generated if no trigger signal appears after approximately 2 sec. This
message goes away when a trigger signal appears.
Backwards Compatibility
Notes
The legacy command:
:TRIGger[:SEQuence]:RFBurst:FSELectivity[:STATe] OFF|ON|0|1
is not supported in the X-Series, as the hardware to do Frequency Selective burst triggers does not
exist in X-Series.
Initial S/W Revision
Prior to A.02.00
Modified at S/W Revision
A.04.00
610
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...