![Keysight E6640A Measurement Applications User & Programmer Reference Download Page 208](http://html1.mh-extra.com/html/keysight/e6640a/e6640a_measurement-applications-user-and-programmer-reference_1961645208.webp)
4 Input/Output Functions
Input/Output
The correction to be used for each bucket is taken from the interpolated correction curve at the center of
the bucket.
Description
Sets an ASCII description field which will be stored in an exported file. Can be displayed in the active
function area by selecting as the active function, if desired to appear in a screen capture.
Key Path
Input/Output, Corrections, Properties
Remote Command
[:SENSe]:CORRection:CSET[1]|2|...|8:DESCription "text"
[:SENSe]:CORRection:CSET[1]|2|...|8:DESCription?
Example
:CORR:CSET1:DESC "11941A Antenna correction"
Notes
45 chars max; may not fit on display if max chars used
Preset
Unaffected by a Preset. Set to empty by Restore Input/Output Defaults
State Saved
Saved in instrument state.
Initial S/W Revision
A.02.00
Modified at S/W Revision
A.14.00
Comment
Sets an ASCII comment field which will be stored in an exported file. Can be displayed in the active
function area by selecting as the active function, if desired to appear in a screen capture.
Key Path
Input/Output, Corrections, Properties
Remote Command
[:SENSe]:CORRection:CSET[1]|2|...|8:COMMent "text"
[:SENSe]:CORRection:CSET[1]|2|...|8:COMMent?
Example
:CORR:CSET1:COMM "this is a comment"
Notes
60 chars max; may not fit on display if max chars used
Preset
Unaffected by Preset. Set to empty by Restore Input/Output Defaults
State Saved
Saved in instrument state
Initial S/W Revision
A.02.00
Modified at S/W Revision
A.14.00
RF Port
Maps one of the sets of corrections to one of the IO ports.
Key Path
Input/Output, Corrections, Properties
Mode
SEQAN
Remote Command
[:SENSe]:CORRection:CSET[1]|2|...|8:RF:PORT RFIN | RFIO1 | RFIO2 |
RFOut | GPSout | GNSSout | RFIO3 | RFIO4
[:SENSe]:CORRection:CSET[1]|2|...|8:RF:PORT?
208
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...