![Keysight E6640A Measurement Applications User & Programmer Reference Download Page 168](http://html1.mh-extra.com/html/keysight/e6640a/e6640a_measurement-applications-user-and-programmer-reference_1961645168.webp)
3 Programming the Test Set
STATus Subsystem
Questionable Integrity Negative Transition
This command determines which bits in the Questionable Integrity Condition register will set the
corresponding bit in the Questionable Integrity Event register when the condition register bit has a negative
transition (1 to 0)
The variable <integer> is the sum of the decimal values of the bits that you want to enable.
Mode
All
Remote Command
:STATus:QUEStionable:INTegrity:NTRansition <integer>
:STATus:QUEStionable:INTegrity:NTRansition?
Example
STAT:QUES:INT:NTR 8 Measurement ‘regained calibration’ Summary will be reported to the Integrity
Summary of the Status Questionable register.
Preset
0
Min
0
Max
32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Integrity Positive Transition
This command determines which bits in the Questionable Integrity Condition register will set the
corresponding bit in the Questionable Integrity Event register when the condition register bit has a positive
transition (0 to 1). The variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Mode
All
Remote Command
:STATus:QUEStionable:INTegrity:PTRansition <integer>
:STATus:QUEStionable:INTegrity:PTRansition?
Example
STAT:QUES:INT:PTR 8 Measurement ‘became uncalibrated’ Summary will be reported to the
Integrity Summary of the Status Questionable register.
Preset
32767
Min
0
Max
32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Integrity Signal Register
"Questionable Integrity Signal Condition" on page 169
168
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...