![Keysight E6640A Measurement Applications User & Programmer Reference Download Page 338](http://html1.mh-extra.com/html/keysight/e6640a/e6640a_measurement-applications-user-and-programmer-reference_1961645338.webp)
7 Trigger Functions
Trigger
Dependencies
Grayed out if in use by Point Trigger in the Source Setup menu.
Forced to Free Run if already selected and Point Trigger is set to External 1.
State Saved
Saved in instrument state
Status Bits/OPC
dependencies
The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping or
Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger event
occurs and all the applicable trigger criteria have been met). A corresponding pop-up message
("Waiting for trigger") is generated if no trigger signal appears after approximately 2 sec. This
message goes away when a trigger signal appears.
Initial S/W Revision
Prior to A.02.00
Trigger Level
Sets the value where the external 1 trigger input will trigger a new sweep/measurement.
Key Path
Trigger, External 1
Remote Command
:TRIGger[:SEQuence]:EXTernal1:LEVel <level>
:TRIGger[:SEQuence]:EXTernal1:LEVel?
Example
TRIG:EXT1:LEV 0.4 V
Couplings
This same level is used for the Ext1 trigger source in the Trigger menu, for the Ext1 selection in the
Periodic Timer sync source (in the Trigger menu and in the Gate Source menu), and also for the Ext1
selection in the Gate Source menu.
Preset
1.2 V
State Saved
Saved in instrument state
Min
–5 V
Max
5 V
Default Unit
V
Backwards
Compatibility SCPI
:TRIGger[:SEQuence]:EXTernal:LEVel
For backward compatibility, the parameter EXTernal is mapped to EXTernal1
Backwards
Compatibility SCPI
:TRIGger[:SEQuence]:FRAMe:EXTernal1:LEVel
Initial S/W Revision
Prior to A.02.00
Trig Slope
Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on a falling
edge.
Key Path
Trigger, External 1
Remote Command
:TRIGger[:SEQuence]:EXTernal1:SLOPe POSitive|NEGative
338
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...