13 Modulation Accuracy (Composite EVM) Measurement
Meas Setup
State Saved
Saved in instrument state.
Range
Std|Left|Right
Initial S/W Revision
Prior to A.02.00
Symbol Boundary (BTS only)
Accesses the menu for the symbol boundary detection modes that are used to make the modulation
accuracy measurement.
•
Auto [SCPI Enum: AUTO] - Sets symbol boundary detection to the automatic mode. Various code
channels are measured and the most appropriate code channel is selected as the reference channel.
Lower level menus enable further selection of various numbers of DPCH channels to make the Modulation
Accuracy Measurement.
•
Test Model 1 - Accesses a menu with further selections of Test Model 1 with 4, 8, 16, 32 or 64 DPCH
channels and with S-CCPCH.
−
Test Model 1 w/4DPCH w/ S-CCPCH [SCPI Enum: TM1D4] - Select this to set the Modulation
Accuracy Measurement to the Test Model 1 with 4 DPCH channels and 1 S-CCPCH channel.
−
Test Model 1 w/8DPCH w/ S-CCPCH [SCPI Enum: TM1D8] - Select this to set the Modulation
Accuracy Measurement to Test Model 1 with 8 DPCH channels and 1 S-CCPCH channel.
−
Test Model 1 w/16 DPCH w/ S-CCPCH [SCPI Enum: TM1D16] - Select this to set the Modulation
Accuracy Measurement to the Test Model 1 with 16 DPCH channels and 1 S-CCPCH channel.
−
Test Model 1 w/32 DPCH w/ S-CCPCH [SCPI Enum: TM1D32] - Select this to set the Modulation
Accuracy Measurement to Test Model 1 with 32 DPCH channels and 1 S-CCPCH channel.
−
Test Model 1 w/64 DPCH w/ S-CCPCH [SCPI Enum: TM1D64] - Select this to set the Modulation
Accuracy Measurement to Test Model 1 with 64 DPCH channels and 1 S-CCPCH channel.
•
Test Model 2 - Accesses a menu with selections of Test Model 2 with S-CCPCH.
−
Test Model 2 w/S-CCPCH [SCPI Enum: TM2SC] - Select this to set the Modulation Accuracy
Measurement to Test Model 2 with 1 S-CCPCH channel.
•
Test Model 3 - Accesses a menu with further selections from Test Model 3 with 4, 8, 16 or 32 DPCH
channels and with S-CCPCH.
−
Test Model 3 w/4 DPCH w/ S-CCPCH [SCPI Enum: TM3D4SC] - Select this to set the Modulation
Accuracy Measurement to Test Model 3 with 4 DPCH channels and 1 S-CCPCH channel.
−
Test Model 3 w/8 DPCH w/ S-CCPCH [SCPI Enum: TM3D8SC] - Select this to set the Modulation
Accuracy Measurement to Test Model 3 with 8 DPCH channels and 1 S-CCPCH channel.
−
Test Model 3 w/16 DPCH w/ S-CCPCH [SCPI Enum: TM3D16SC] - Select this to set the Modulation
Accuracy Measurement to Test Model 3 with 16 DPCH channels and 1 S-CCPCH channel.
−
Test Model 3 w/32 DPCH w/ S-CCPCH [SCPI Enum: TM3D32SC] - Select this to set the Modulation
Accuracy Measurement to Test Model 3 with 32 DPCH channels and 1 S-CCPCH channel.
W-CDMA & HSPA Measurement Applications Reference
2079
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...